| Model Number |
MMT-723LNAS |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm and it kept loading.The customer reported that the screen went off and it started counting.The customer was able to clear the alarm but was not able to complete rewind.No harm requiring medical intervention was reported.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test.However device alarmed a33 error during basic occlusion test due to loose or protruded drive support disk.Unable to perform occlusion test, prime or a33 test, excessive no delivery test.
|
| |
|