• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH 40 SILICONE PIP IMPLANT (STERILE PACKED); PROSTHESIS, FINGER, CONSTRAINED, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH 40 SILICONE PIP IMPLANT (STERILE PACKED); PROSTHESIS, FINGER, CONSTRAINED, POLYMER Back to Search Results
Model Number PIP-40
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348)
Date of Event 01/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that the patient's hand (knuckle) was revised due to the pip silicone implant breaking.A 40 silicone pip implant was revised to another 40 silicone pip implant.
 
Additional Manufacturer Narrative
The reported event could only partially be confirmed, since the device was returned, but no x-rays were provided.More detailed information about the complaint event (x-rays, operative surgery.) must be available in order to determine the root cause of the complaint event.The device inspection revealed the following: the product was returned and shows severe signs of wear (multiple scratches are noticed).Moreover, the implant is partially cut in two.Unfortunately, the analysis of the returned device alone is not sufficient to allow us to draw conclusions on this case.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
 
Event or Problem Description
It was reported that the patient's hand (knuckle) was revised due to the pip silicone implant breaking.A 40 silicone pip implant was revised to another 40 silicone pip implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
40 SILICONE PIP IMPLANT (STERILE PACKED)
Common Device Name
PROSTHESIS, FINGER, CONSTRAINED, POLYMER
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key9659480
Report Number0008031020-2020-00253
Device Sequence Number19670656
Product Code KYJ
UDI-Device Identifier00886385021577
UDI-Public00886385021577
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K931588
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPIP-40
Device Catalogue NumberPIP40
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2020
Initial Date Received by Manufacturer 01/06/2020
Supplement Date Received by Manufacturer02/25/2020
Initial Report FDA Received Date02/03/2020
Supplement Report FDA Received Date03/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-