| Model Number |
PIP-40 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Injury (2348)
|
| Date of Event |
01/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's hand (knuckle) was revised due to the pip silicone implant breaking.A 40 silicone pip implant was revised to another 40 silicone pip implant.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could only partially be confirmed, since the device was returned, but no x-rays were provided.More detailed information about the complaint event (x-rays, operative surgery.) must be available in order to determine the root cause of the complaint event.The device inspection revealed the following: the product was returned and shows severe signs of wear (multiple scratches are noticed).Moreover, the implant is partially cut in two.Unfortunately, the analysis of the returned device alone is not sufficient to allow us to draw conclusions on this case.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's hand (knuckle) was revised due to the pip silicone implant breaking.A 40 silicone pip implant was revised to another 40 silicone pip implant.
|
| |
|