| Catalog Number |
SHORT-HD |
| Medical Device Problem Codes |
Vibration (1674); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The assignable root cause was traced to component failure due to faulty parts.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported from (b)(6) that during service and repair/pre-testing, it was observed that the attachment device bearing was worn and the triangle symbol was missing.The device also failed pretests for preload, vibration and temperature assessment.The event was not related to surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|