| Model Number |
8888160556 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported that the nurse practitioner was adjusting placement of an umbilical venous catheter and noticed that there was a drop of iv fluid escaping from the side of the umbilical catheter.The point where the fluid was leaking was close to where the line was sutured; the nurse practitioner noticed that the suture was quite tight.
|
| |
|
Additional Manufacturer Narrative
|
|
H 3: evaluation summary: because no lot number was available, a manufacturing device history record (dhr) review or product/process changes review for the involved lot number could not be performed.However, all dhrs are reviewed for accuracy prior to product release.One used catheter was received at the manufacturing site for analysis and investigation.Visual inspection of the sample presented signs of use.During the evaluation the catheter showed a leak below the strain relief.An underwater test was performed and a leak below the strain relief could be identified on the catheter.The reported condition has been confirmed.The most probable root cause can be considered as the catheter was most likely damage during use due to a sharp instrument or manipulation.It must be noted that in-process controls, such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling, are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
|
| |
|
Search Alerts/Recalls
|