| Model Number |
71371-75 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.The device manufacturer date for the reported meter is unknown.The date entered is the date of the event.Note: the meter is designed to report readings of 20 mg/dl to 500 mg/dl.It should be noted that this meter does not give numeric value for readings less than 20 mg/dl.A "lo" display message indicates a reading less than 20 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of "lo" (less than 20 mg/dl), and 113 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The dhrs (device history review) for the freestyle precision neo meter was reviewed.The dhrs showed the freestyle precision neo meter passed all tests prior to release.The dhr for the precision strips was reviewed, and the dhr showed the precision strips passed all tests prior to release.Retain testing was performed and all units performed within specification.If the product is returned, the case will be re-opened, and a physical investigation will be performed.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of "lo" (less than 20 mg/dl), and 113 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
|
| |
|