BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2R8537
Device Problems Filling Problem (1233); Improper Flow or Infusion (2954)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified quantity of clearlink system non-dehp continu-flo solution sets under infused and caused air in line alarms while in use with infusion pumps.This was further described as; ¿the pump delivered the drug over the specified amount of time¿ and ¿air in line alarm went off¿ with use of (non-baxter) closed system drug transfer device (cstd).The customer stated, they ¿were able to see actual air bubbles in the line when the air in line alarm occurs¿.This was identified during patient infusions.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerrfield IL
MDR Report Key9662465
MDR Text Key178118366
Report Number1416980-2020-00373
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
PMA/PMN Number
K180739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2R8537
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/07/2020
Initial Date FDA Received02/03/2020
Supplement Dates Manufacturer Received02/19/2020
Supplement Dates FDA Received03/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
BAXTER SPECTRUM IQ PUMP; PHASEAL CSTD; UNSPECIFIED DRUG
-
-