• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I FULL OSSEOTITE® IMPLANT 5 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I FULL OSSEOTITE® IMPLANT 5 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number FOS510
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 11/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).
 
Event or Problem Description
It was reported implant removed due to perimplantitis.
 
Additional Manufacturer Narrative
Peri-implantitis is an infectious disease that causes inflammation of the gum and the bone structure around a dental implant.Chronic inflammation causes bone loss, which can lead to implant failure thus its removal.Investigations performed for hundreds of events where either of these conditions, infection or bone loss, or both, have been reported, have concluded that the likely cause(s) for the implant failure in relation to these conditions are external factors.Those include, medical conditions (e.G., diabetes, bruxism, etc.), patient habits (e.G., smoking, bad oral hygiene routine) and user error (e.G., surgical technique).There has not been any instance where either infection and/or bone loss, and when right conditions prevail and led to peri-implantitis whether reported or not along with the other two, have resulted from a manufacturing or design defect.Furthermore, the probability of manufacturing or design defects that might lead to infection and result in bone loss occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming product.The analysis and result of investigations and the analysis of probability previously described are contained in the summary investigation reports performed for bone loss and infection.Additionally, all device history record reviews verified that each implant was sterilized per procedure for every device.All complaint data used for the summary investigation was found to be conforming and did not meet capa/hhe/d/ie escalation.Therefore, there were no complaints which confirmed a manufacturing or design related issue that did or could cause or contribute to the reported event.Dhr review was completed for the subject lot number.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimmer biomet control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL OSSEOTITE® IMPLANT 5 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key9664098
Report Number0001038806-2020-00255
Device Sequence Number19533115
Product Code DZE
UDI-Device Identifier00844868003724
UDI-Public(01)00844868003724(17)121205(10)764682(241)FOS510
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K063286
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2019
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/05/2012
Device Model NumberFOS510
Device Catalogue NumberFOS510
Device Lot Number764682
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/14/2020
Supplement Date Received by Manufacturer04/29/2020
Initial Report FDA Received Date02/04/2020
Supplement Report FDA Received Date04/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient Weight70
-
-