| Model Number |
10603 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.00 x 20 synergy ii drug-eluting stent selected for use.However, while inserting the device into the guide, the stent catheter shaft broke.The procedure was completed with another of same device.There were no patient complications nor injuries reported and the patient was fine.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 3.00 x 20 synergy ii drug-eluting stent selected for use.However, while inserting the device into the guide, the stent catheter shaft broke.The procedure was completed with another of same device.There were no patient complications nor injuries reported and the patient was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device is a combination product.Device evaluated by mfr: synergy ii us mr 3.00 x 20 mm stent delivery system was returned for analysis attached to a haemostatic valve.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks and a break 23.5cm distal to the distal end of the strain relief.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Search Alerts/Recalls
|