• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN GIENTRI PORT WITH ENFIT CONNEC; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN GIENTRI PORT WITH ENFIT CONNEC; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 771111E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
 
Event or Problem Description
The customer reported that the sump bifurcated connector feeding device was leaking at the dial portion of the port.There was no harm to the patient.
 
Additional Manufacturer Narrative
H 3 evaluation summary: a device history record (dhr) review could not be completed because the lot number was unavailable.The sample was received for evaluation.A leak in the received sample was not confirmed, however a missing component was detected in the anti-reflex valve (arv).The reported condition could not be confirmed.The root cause for the reported condition cannot be specifically identified; therefore, corrective action will be limited to manufacturing awareness at this time.The current process is running according to product specifications, meeting quality acceptance criteria.The manufacturing site will continue monitoring the process for any adverse trends that require immediate attention.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GIENTRI PORT WITH ENFIT CONNEC
Common Device Name
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
COVIDIEN
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
MDR Report Key9664763
Report Number9612030-2020-02405
Device Sequence Number2756740
Product Code PIF
UDI-Device Identifier10884521582750
UDI-Public10884521582750
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number771111E
Device Catalogue Number771111E
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 01/23/2020
Supplement Date Received by Manufacturer01/23/2020
Initial Report FDA Received Date02/04/2020
Supplement Report FDA Received Date04/02/2020
Patient Sequence Number1
-
-