| Model Number |
AH20S |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
12/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Weight unknown / not provided.Pma/510(k): k011028 and k013227.
|
| |
|
Event or Problem Description
|
|
It was reported that screw fractured.Implant is in good condition.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information.H6: method code was updated to 3331 and 4109.H6: results code was updated to 3252.H6: conclusions code was updated to 4307.The reported device was received for evaluation.The reported condition of a fractured screw was confirmed.Visual evaluation of the as returned product identified significant signs of wear around the threads and fracture at the screw head.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.A device history record (dhr) review was completed for the reported device lot.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformance, which could have caused or contributed to the reported event was noted as part of the dhr.A complaint history review was completed for the reported device lot for similar event and no other complaint was identified.A definitive root cause for the reported event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|