• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HARH36
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/07/2020
Event Type  Injury  
Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Visual inspection revealed evidence of clinical use.The distal end of the blade was fractured and separated from the rest of the device.The device showed signs of erosion/gouges in the blade coating.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: jaws/blade subassembly damage incidental and prolonged activation against solid surfaces, such as bone or plastic the instructions for use (ifu) state: use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.Do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.Avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.Note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.Note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
 
Event Description
It was reported the blade of the harmonic scalpel tip broke off into the patient.The x-ray machine was brought in to ensure all metal was removed.The piece was retrieved completely from the patient through the same incision.There was no patient injury and extended procedure time reported was minimal at 20-30 minutes.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
Manufacturer Contact
marissa richmond
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key9665020
MDR Text Key182706046
Report Number0001056128-2020-00007
Device Sequence Number1
Product Code NLQ
UDI-Device Identifier07613327346251
UDI-Public07613327346251
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161693
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/26/2022
Device Model NumberHARH36
Device Catalogue NumberHARH36RR
Device Lot Number10644261
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-