| Model Number |
401736 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An orsiro was selected for treatment.Before introducing into patient, it was noticed that the stent was deformed.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was available for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Based on the conducted investigations of the device being subject to this complaint, no manufacturing or material related root cause could be determined.
|
| |
|
Search Alerts/Recalls
|