| Model Number |
MMT-723NAP |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/31/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm.The customer stated the insulin pump not dropped or bumped.Customer reported that the drive support cap was sticking out, they were able to push it back in.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received a33 alarm during the basic occlusion test due to loose or protruded or flush drive support disk.Unable to perform prime or a33 test, excessive no delivery test and occlusion test due to a33 alarm.Device passed the displacement test.The test reservoir p-cap was able to lock in place.
|
| |
|