• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 20MM; ORTHOPEDIC BONE SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 20MM; ORTHOPEDIC BONE SCREW Back to Search Results
Catalog Number 010000997
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: g7 screw 6.5mm x 20mm: cat#010000997, lot# 6591649.G7 screw 6.5mm x 20mm: cat#010000997, lot# 6591649.G7 screw 6.5mm x 20mm: cat#010000997, lot# 6591649.G7 screw 6.5mm x 20mm: cat#010000997, lot#6591649.G7 screw 6.5mm x 35mm: cat#010001000, lot#6586581.G7 screw 6.5mm x 20mm: cat#010000997, lot#6591649.G7 screw 6.5mm x 20mm: cat#010000997, lot#6591649.G7 screw 6.5mm x 35mm: cat#010001000, lot#6586581.G7 screw 6.5mm x 35mm: cat#010001000, lot#6586581.Foreign country: (b)(6).Visual inspection of the returned units for lot # 6591649 found that 3 of 7 units meet the specifications and 4 of 7 units are below the 6mm specification.The reported issue is confirmed for 4 of 7 units.Visual inspection of the returned units for lot # 6586581 found that all 3 units did not meet the seal width specifications.The reported issue is confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history identified additional similar complaints for the reported item(s) and no additional complaints for the reported part and lot combination(s).The likely condition of the product when leaving zimmer biomet was not conforming to specifications.The root cause of the reported issue is attributed to operator error during the manufacturing process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-00505, 0001825034-2020-00506, 0001825034-2020-00508, 0001825034-2020-00510, 0001825034-2020-00511, 0001825034-2020-00512.
 
Event or Problem Description
It was reported through incoming inspection at the warehouse that the seal width was less than 1/4 inch (6.35mm).No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 SCREW 6.5MM X 20MM
Common Device Name
ORTHOPEDIC BONE SCREW
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9670864
Report Number0001825034-2020-00509
Device Sequence Number379574
Product Code PBI
UDI-Device Identifier00880304527454
UDI-Public(01)00880304527454(17)290714(10)6591649
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number010000997
Device Lot Number6591649
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2019
Initial Report FDA Received Date02/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-