Avanos medical inc.Received a single report that referenced six different incidences, which were associated with separate units, involving different patients.This is the sixth of six reports.Refer to 3011270181-2020-00018 for the first report.Refer to 3011270181-2020-00019 for the second report.Refer to 3011270181-2020-00020 for the third report.Refer to 3011270181-2020-00021 for the fourth report.Refer to 3011270181-2020-00022 for the fifth report.It was reported that over the past 21-30 days the user facility experienced multiple incidents of the ett [endotracheal tube] developing tears in the cuff.The tears occurred during use with multiple doctors who were all very familiar with using the product.No patient injuries; however, all incidents required patient to be reintubated.Additional information received from the customer on 28-jan-2020 indicated "we had a total of 6 ett changes due to cuff failure.It was not the same patient each time.The tears were noted in the cuff itself upon removal of the ett, but was not noted prior to inspection.".
|