| Model Number |
N/A |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Report source: (b)(6).Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned product identified white debris inside the sterile packaging from the foam packaging.Dhr was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to transit damage causing the foam to shed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during warehouse inspection, debris was identified within the sterile package.There was no patient involvement.Attempts have been made and additional information on the event is unavailable at this time.
|
| |
|