MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Failure to Interrogate (1332); Overheating of Device (1437); Communication or Transmission Problem (2896); Connection Problem (2900)
|
Patient Problem
Burn(s) (1757)
|
Event Date 01/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 97755, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(4), udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received regarding a patient with an implantable neurostimulator (ins), and it was reported that the recharge telemetry module wire is bent where it goes into the paddle.The patient stated that the paddle got hot and would burn her.The patient stated she was burned by the recharge telemetry module but it did not leave a mark.The patient was sent a new controller for an unrelated issue, but it would not find the recharge telemetry module which would not find the ins.Patient was issued a new recharge telemetry module.There were no reported complications and no further complications were expected.
|
|
Manufacturer Narrative
|
Product id 97755, serial# (b)(4).Product type recharger; analysis found that the recharger antenna was damaged and the coating at entrance of donut end was splitting open.If information is provided in the future, a supplemental report will be issued.
|
|