MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PARTIAL TIBIAL CEMENTED SIZE F LEFT MEDIAL; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PARTIAL TIBIAL CEMENTED SIZE F LEFT MEDIAL; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Patient Device Interaction Problem (4001); Migration (4003)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994)
Date of Event 01/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Report source: event occurred in (b)(6).
 
Event or Problem Description
It was reported that approximately 1 month post implantation, the patient experienced pain and was found to have a tibial bone fracture.The patient has indicated a possible revision, but one has not been confirmed yet.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
No devices or photos were received; therefore, the condition of the components is unknown.The dhr was reviewed and no discrepancies relevant to the reported event were found.X-rays evaluation report provided by third party hcp states that there is evidence of a vertically oriented medial tibial plateau fracture and associated subsidence of the tibial component.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARTIAL TIBIAL CEMENTED SIZE F LEFT MEDIAL
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9672279
Report Number0001825034-2020-00490
Device Sequence Number19563955
Product Code HSX
UDI-Device Identifier00880304812703
UDI-Public(01)00880304812703
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42538000601
Device Lot Number64271273
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/31/2020
Supplement Date Received by Manufacturer03/09/2020
Initial Report FDA Received Date02/05/2020
Supplement Report FDA Received Date03/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Weight60
-
-