| Model Number |
SONATA |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user lost hearing benefit with the device about 9 months ago.The user was re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Damage to the active electrode, likely caused by minute device mobility, was determined to have led to device failure over time.Other damages are most likely related due to explantation surgery.The problems given in the recipient report appear to match the damage found.This is the final report.
|
| |
|
Event or Problem Description
|
|
The user lost hearing benefit with the device about 9 months ago.The user was re-implanted.
|
| |
|