On (b)(6) 2020, abbott point of care was contacted by a customer regarding i-stat ec8+ cartridges that yielded a suspected discrepant potassium result of 7 on a patient.There was no patient information available at the time of this report.Method: i-stat, k results: 7 mmol/l.Lab, 4 mmol/l.Collect, test times, and cartridge lot information was not provided.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
|
Apoc incident # (b)(4).Product lot# information that was used during the event was unknown on the initial call.On 02-mar-2020 an abbott support specialist stated that the following chem8+ lots were shipped to add france in january 2020.Product investigated per shipping records.(d4) lot #: (d4) expiration date: (h4) device manufacture date: h19326a, 05/20/2020, 11/22/2019.H19333b, 05/27/2020, 11/29/2019.H19334b, 05/28/2020, 11/30/2019.H19339a, 06/02/2020, 12/05/2019.H19342b, 06/05/2020, 12/08/2019.The investigation was completed on 03/31/2020.A review of the device history records (dhrs) confirmed the cartridge lots met finished goods (fg) release criteria.Retained testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ae (product complaint level 2 and level 3 investigation procedure).No deficiency has been determined for chem8+ lots h19326a, h19333b, h19334b, h19339a and h19342b.
|