• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AD-TECH MEDICAL INSTRUMENT CORP. CRANIAL DRILL BITS AND ACCESSORIES; DRILL SLEEVE GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AD-TECH MEDICAL INSTRUMENT CORP. CRANIAL DRILL BITS AND ACCESSORIES; DRILL SLEEVE GUIDE Back to Search Results
Model Number DSG-6.3-090-2.4N
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mdr 2183456-2019-00016 has been filed past the 30-day timeframe.Through the 2019 fda inspection, it was identified that while ad-tech was evaluating reportability for actual events in which death or serious injury had occurred, ad-tech had failed to submit events in which a malfunction was reported and could result in a serious injury if the event were to recur.As part of the investigation, a two (2) year retrospective review of complaints was performed to determine which complaints required submission of an mdr.Mdr-2183456-2019-00016 was submitted as a conservative measure due to the recalls tied to this issue.
 
Event or Problem Description
On (b)(6) 2018 a user facility reported issues with an ad-tech drill bit and sleeve guide in an seeg case they had that day.The drill sleeve blocks were fused to the drill sleeve guide.The customer used a backup and the case proceeded as planned without any other issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIAL DRILL BITS AND ACCESSORIES
Common Device Name
DRILL SLEEVE GUIDE
Manufacturer (Section D)
AD-TECH MEDICAL INSTRUMENT CORP.
400 west oakview parkway
oak creek WI 53154
MDR Report Key9673579
Report Number2183456-2019-00016
Device Sequence Number19560244
Product Code HBE
UDI-Device Identifier00841823106520
UDI-Public(01)00841823106520(17)200601(10)0122065819260
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K170442
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2020
Device Model NumberDSG-6.3-090-2.4N
Device Catalogue NumberDSG-6.3-090-2.4N
Device Lot Number819260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/29/2018
Initial Report FDA Received Date02/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2183456-08212019-R-004
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-