| Model Number |
DDK2-2.4-30X |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mdr 2183456-2019-00034 has been filed past the 30-day timeframe.Through the 2019 fda inspection, it was identified that while ad-tech was evaluating reportability for actual events in which death or serious injury had occurred, ad-tech had failed to submit events in which a malfunction was reported and could result in a serious injury if the event were to recur.As part of the investigation, a two (2) year retrospective review of complaints was performed to determine which complaints required submission of an mdr.Mdr-2183456-2019-00034 was submitted as a conservative measure due to the recalls tied to this issue.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, an ad-tech clinical specialist was notified about an issue that occurred when ad-tech's drill bit locked into the guide.The case proceeded with a backup drill and there was no report of patient harm.
|
| |
|
Search Alerts/Recalls
|