| Model Number |
MMT-523LNAB |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Power Problem (3010); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump hade code error.Customer stated that the insulin pump had rejected brand new battery.Customer reported that the insulin pump had no delivery alerts.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with all operating currents within specification and passed functional testing including the rewind, a21 error test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.No excessive no delivery or occlusion alarm noted.No unexpected failed battery alarm anomalies noted.No unexpected e/a alarm anomalies noted.This mdr related to the (b)(4) manufacturing site has been assigned a medwatch number from the medtronic minimed (b)(4) site, per variance 5.
|
| |
|