| Model Number |
402804 |
| Medical Device Problem Codes |
Material Frayed (1262); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission. .
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, the catheter would not deflect as intended while mapping.The catheter was removed from the patient and upon visual inspection, the insulation of the catheter shaft had split.The catheter was replaced and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported shaft damage was confirmed.Visual inspection showed a fracture in the shaft at the spiral loop/shaft transition.Further investigation concluded that the fracture in the shaft was due to a weak bonding at the spiral loop/shaft transition.
|
| |
|