• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 5.5PLC
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Extubate (2402)
Date of Event 01/22/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during use, a cuff check was performed before replacing cannula.After 3 hours of replacement, there was a cuff leakage and was replaced.
 
Additional Manufacturer Narrative
Additional information: evaluation summary: one sample of device was received for evaluation.An inflation/deflation test was performed and it was observed the cuff deflated after 15 minutes, the sample was submerged into a container filled with water and it was observed the pilot balloon leaked.The reported event was confirmed.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
According to the reporter, during use, a cuff check was performed before replacing cannula.After 3 hours of replacement, there was a cuff leakage and was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
9560 jose rodriguez drive
el paso TX 79927
MDR Report Key9676542
Report Number2936999-2020-00099
Device Sequence Number19625329
Product Code JOH
UDI-Device Identifier20884522029046
UDI-Public20884522029046
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K955680
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5.5PLC
Device Catalogue Number5.5PLC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2020
Initial Date Received by Manufacturer 01/24/2020
Supplement Date Received by Manufacturer03/17/2020
Initial Report FDA Received Date02/06/2020
Supplement Report FDA Received Date04/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-