This system was explanted due to erosion.There is currently no indication that a new system has been implanted.Lead broke during extraction and was partially explanted, but is mostly still in the body.Should additional information become available, this file will be updated.
|
|
An erosion was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the erosion was not device related.
|