• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIARADXT
Medical Device Problem Codes Loss of or Failure to Bond (1068); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fever (1858); Pain (1994); No Code Available (3191)
Date of Event 01/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, post hemicolectomy, patient presented some pain and fever.It was stated that the patient went back to surgery and extend hospital stay.
 
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the reload had a full staple complement.Functionally the reload was loaded into a pmv representative instrument and was applied to test media with proper staple placement and media transection.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, everything appeared to be okay at the time of the anastomosis of an open hemicolectomy, but the surgeon had doubts due to the lower resection and decided to proceed with a preventive or protective ileostomy.However, the patient developed septicemia due to failure of the anastomosis and presented some pain and fever nine days following the hemicolectomy procedure.The surgeon sent the patient back to surgery for exploratory laparotomy and found the mishap of the staple line was open in the anastomosis.The patient has been treated with cefotaxime and metronidazole.This resulted to 28 days extended hospital stay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key9677849
Report Number1219930-2020-00615
Device Sequence Number14052522
Product Code GDW
UDI-Device Identifier10884521141742
UDI-Public10884521141742
Combination Product (Y/N)N
Initial Reporter CountryCS
PMA/510(K) Number
K102291
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberEGIARADXT
Device Catalogue NumberEGIARADXT
Device Lot NumberN8M0211Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2020
Initial Date Received by Manufacturer 01/24/2020
Supplement Date Received by Manufacturer04/09/2020
Initial Report FDA Received Date02/06/2020
Supplement Report FDA Received Date04/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight96
-
-