| Model Number |
367008 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A pantera leo balloon catheter was chosen for treatment of a severely calcified lesion.During inflation the balloon ruptured at 18 atm.Thus another balloon was used for the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinal over a length of about 1.5 mm.Microscopic inspection of the balloon showed several scratches in close vicinity of the tear.It seems likely that the tear in the balloon was caused by a hard, sharp-edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause was determined.The root cause is most likely related to the patients anatomy (i.E.Calcified lesion).
|
| |
|
Search Alerts/Recalls
|