| Model Number |
7508 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Used the first day of the month of aware date as no event date was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.A 1.50mm x 20mm apex flex balloon catheter was advanced for dilatation.However, the balloon was unable to be advanced nor pull back from the wire.Both balloon and guidewire were removed from patient's body as a unit.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.A 1.50mm x 20mm apex flex balloon catheter was advanced for dilatation.However, the balloon was unable to be advanced nor pull back from the wire.Both balloon and guidewire were removed from patient's body as a unit.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Used the first day of the month of aware date as no event date was provided.Device evaluated by mfr.:returned product consisted of an apex balloon catheter with a non-bsc 0.014 inch guidewire in the lumen.The hub, shafts, tip, and balloon were microscopically and visually examined.There was contrast in the inflation lumen.The balloon was tightly folded.Inspection of the device revealed that the wire lumen shaft was buckled in the hub of the device with the wire lumen wavy for 18cm distal of the hub.The outer shaft was stretched 97cm distal of the hub to the proximal end of the balloon.A measurement of the shaft was not able to be performed with the shaft being stretched.The guidewire used in the procedure was received stuck inside the complaint device.An attempt to remove the guidewire was unsuccessful, as it was stuck on wire due to the buckled and stretched wire lumen shaft.The damage to the wire lumen shaft is consistent to damage caused by the guidewire during advancement or removal.
|
| |
|