STRYKER ORTHOPAEDICS-MAHWAH SECUR-FIT MAX 127 HIP STEM #9; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 6052-0935S
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 01/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.
 
Event or Problem Description
Dr.Revised a right total hip that was originally done on (b)(6) 2019.Patient had dislocated multiple times.Dr.Felt the cup was placed too vertically and the stem was undersized.He revised the entire hip to a trident ii multi-hole cup, mdm and omnifit hfx press fit stem.Update 13/january/2020: spoke to rep.Surgeon also felt that the implanted stem was undersized for the patient's anatomy and was revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECUR-FIT MAX 127 HIP STEM #9
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9678665
Report Number0002249697-2020-00241
Device Sequence Number16268979
Product Code LZO
UDI-Device Identifier07613327021127
UDI-Public07613327021127
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K051738
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number6052-0935S
Device Catalogue Number6052-0935S
Device Lot NumberJN0J0E
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2020
Initial Report FDA Received Date02/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
-
-