Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
Patient information was not provided.Serial number is unknown.This information will be provided in a supplemental report if made available.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Through media monitoring livanova learned of an infection of a patient with m.Chimaera.In the paper article it is stated that the patient had started to feel ill during 2019 showing increasing weakness, a high weight loss, some fever.The infected patient, hospitalized in the last few months at (b)(6), has now been discharged to continue antibiotic therapy at home.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: further information has been requested to the hospital in order to confirm/deny the news.Despite the attempts, the hospitals did not provide any feedback to livanova.Thus, no further investigation is possible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
MDR Report Key9680292
MDR Text Key188837231
Report Number9611109-2020-00062
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2020
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received03/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-