• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1000 MED-EL CONCERT PIN
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user's hearing performance with the device is affected.It is reported that the patient did not present any responses in the free field audiometry (not able to detect loud sounds).
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
 
Event or Problem Description
The user's hearing performance with the device is affected.It is reported that the patient did not present any responses in the free field audiometry (not able to detect loud sounds).Re-implantation was scheduled to take place in (b)(6) 2020.However, due to covid-19 the surgery has been postponed.
 
Additional Manufacturer Narrative
Conclusion: damage to the active electrode, as might be caused by an external mechanical impact, was determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Event or Problem Description
The user's hearing performance with the device is affected.It is reported that the recipient is not able to detect loud sounds.The recipient was re-implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key9680465
Report Number9710014-2020-00085
Device Sequence Number13578572
Product Code MCM
UDI-Device Identifier09008737062736
UDI-Public(01)09008737062736
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMI1000 MED-EL CONCERT PIN
Device Catalogue Number07679
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/30/2020
Supplement Date Received by Manufacturer01/30/2020
01/30/2020
Initial Report FDA Received Date02/07/2020
Supplement Report FDA Received Date05/01/2020
10/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 MO
-
-