According to the reporter, during a veterinary case, the device needle detached from two separate packages of suture as it was being withdrawn from the packaging.The surgeon then used another device to resolve the issue in order to complete the case.There was no patient injury.
|
|
Additional information: a1, a2, a3, a4, b5, d10, g4, h3, h6 h3 evaluation summary: post market vigilance (pmv) led an evaluation of thirty devices, two opened by the account without packaging and twenty-eight sealed.The visual inspection of the sample noted two sutures and two needles separated.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.Further microscopic inspection noted that the swage was filled of suture transfer material.A needle attachment test was performed on the sealed samples, and the results met the specifications for this product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
According to the reporter, during a veterinary case, the device needle detached from 2 separate packages of suture with the same lot number, as it was being withdrawn from the packaging.First happened while the surgeon taking the needle out, second happened on the first pass through the skin.The surgeon then used another device to resolve the issue in order to complete the case.There was no patient injury.
|