The reported event, of pocket heating was not confirmed.The alleged complaint for pocket heating could not be tested, due to insufficient battery voltage.Since cutting the can open to supply, sufficient battery voltage would invalidate the pocket heating test.Destructive analysis was not performed on the device.Testing of the ipg did not identify any anomalies, that would contribute to the alleged complaint of pocket heating.The ipg performed normally, during testing.And a review of the ipg logs indicated, normal current draw over the life of the ipg.The intense heating and burning allegation, that the patient experienced in late (b)(6) of 2019, postdates the ipg end of service (eos).Ipg stimulation turned off, due to eos on (b)(6) 2019.Since ipg stimulation was turned off, due to eos.The alleged complaint of pocket heating could not have been caused by the ipg.
|