MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3507125BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: rupture.Concomitant products: mentor memorygel breast implant 125cc, catalog # 3507125bc, serial # (b)(4).Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old caucasian female patient underwent a breast augmentation revision with mentor memorygel breast implant 125cc and experienced a rupture on the left side post-operatively, which was confirmed by a physician.As a result, the patient underwent explantation on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
On march 17, 2020, mentor completed evaluation of the device.Device evaluation summary: during visual inspection of the device, a crease was noted on the posterior aspect.Additionally, a tear was observed within the crease measuring approximately 0.2 cm.The evaluation determined that the rupture is consistent with a crease fold rupture.These kind of findings are consistent with a crease/fold failure which is a known inherent risk associated with the use of gel-filled mammary prostheses and may be the result of one or more of the following contributing factors: continuous and sustained stresses to the device - such as too small a breast pocket and folding or wrinkling of the shell in the breast pocket.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2020, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|