Model Number 3507125BC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: rupture.Concomitant products: mentor memorygel breast implant 125cc, catalog # 3507125bc, serial # (b)(4).Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old caucasian female patient underwent a breast augmentation revision with mentor memorygel breast implant 125cc and experienced a rupture on the left side post-operatively, which was confirmed by a physician.As a result, the patient underwent explantation on (b)(6) 2020.
 
Additional Manufacturer Narrative
On march 17, 2020, mentor completed evaluation of the device.Device evaluation summary: during visual inspection of the device, a crease was noted on the posterior aspect.Additionally, a tear was observed within the crease measuring approximately 0.2 cm.The evaluation determined that the rupture is consistent with a crease fold rupture.These kind of findings are consistent with a crease/fold failure which is a known inherent risk associated with the use of gel-filled mammary prostheses and may be the result of one or more of the following contributing factors: continuous and sustained stresses to the device - such as too small a breast pocket and folding or wrinkling of the shell in the breast pocket.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.(b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2020, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key9682651
Report Number1645337-2020-02020
Device Sequence Number11551968
Product Code FTR
UDI-Device Identifier00081317000020
UDI-Public00081317000020
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2015
Device Model Number3507125BC
Device Catalogue Number3507125BC
Device Lot Number6435408
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2020
Initial Date Received by Manufacturer 01/17/2020
Supplement Date Received by Manufacturer02/19/2020
03/17/2020
Initial Report FDA Received Date02/07/2020
Supplement Report FDA Received Date03/11/2020
03/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
-
-