Model Number 51-405-42-71
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 12/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Reference exemption number e2017029.
 
Event or Problem Description
It was reported the patient did not reach full distraction.The distractor was removed.
 
Additional Manufacturer Narrative
An investigation was performed in the lab.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The investigation results conclude that the root cause for could not be determined.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTRACTION
Common Device Name
DISTRACTION
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
MDR Report Key9682992
Report Number9610905-2020-00004
Device Sequence Number3445909
Product Code PBJ
UDI-Device Identifier00888118114310
UDI-Public(01)00888118114310
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K163315
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/15/2020
Report Date (Section F) 01/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number51-405-42-71
Device Lot Number33306048
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/24/2020
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/10/2020
Event Location Hospital
Date Report to Manufacturer01/14/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2020
Supplement Date Received by Manufacturer06/10/2020
Initial Report FDA Received Date02/07/2020
Supplement Report FDA Received Date07/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age6 MO