| Model Number |
174021 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the device jammed upon firing.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: post market vigilance (pmv) led an evaluation of one device.The e-piece was partially disengaged from the device on one side.There was evidence of weld on the device.The loading unit was received preloaded on the instrument with three tacks remaining.The instrument was applied to the appropriate test media.The remaining three tacks deployed improperly and did not seat properly due to the disengaged e-piece.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur as a result of excessive manipulation of the device.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|