• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO MULTIFIRE VERSATACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO MULTIFIRE VERSATACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 174021
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the device jammed upon firing.There was no patient injury.
 
Additional Manufacturer Narrative
Investigation summary: post market vigilance (pmv) led an evaluation of one device.The e-piece was partially disengaged from the device on one side.There was evidence of weld on the device.The loading unit was received preloaded on the instrument with three tacks remaining.The instrument was applied to the appropriate test media.The remaining three tacks deployed improperly and did not seat properly due to the disengaged e-piece.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur as a result of excessive manipulation of the device.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE VERSATACK
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key9684062
Report Number2647580-2020-00524
Device Sequence Number2837270
Product Code GCJ
UDI-Device Identifier10884521070561
UDI-Public10884521070561
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K912097
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 03/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number174021
Device Catalogue Number174021
Device Lot NumberP8A1073X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2020
Initial Date Received by Manufacturer 01/17/2020
Supplement Date Received by Manufacturer02/20/2020
Initial Report FDA Received Date02/07/2020
Supplement Report FDA Received Date03/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-