| Model Number |
APTIO AUTOMATION |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A siemens customer service engineer (cse) observed sparks while replacing an electrical relay for the bulk input module (bim) on an aptio automation system.The aptio automation system was under the control of a siemens customer service engineer (cse) at the time of the event.There was no report of smoke or flame.There was no delay in processing samples.The spark was caused by the installation of an incorrect part number relay in the bim.The cse obtained the correct part and installed it without issue.The instrument is performing within specifications.No further evaluation of this device is needed.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020 a siemens customer service engineer (cse) observed sparks while replacing an electrical relay for the bulk input module (bim) on an aptio automation system.There was no report of smoke or flame.There was no delay in processing samples on the aptio automation system.There are no known reports of patient intervention, or adverse health consequences due to the observed sparks.
|
| |
|