CYBERONICS - HOUSTON PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 3000
Medical Device Problem Codes Computer Software Problem (1112); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/19/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported via a programming history database periodic review that the patient had high impedance on two separate programmers.The high impedance was observed with (b)(4) software.For model 102/102r generators programmed to output currents > 1 ma , it was observed that system diagnostics using m3000 (b)(4) and m250 software would display ok lead impedance while system diagnostics using m3000 (b)(4) software would display a false high impedance.No other relevant information has been received to date.Mfr.Report # 1644487-2020-00213 captures the high impedance seen with programmer 1.Mfr.Report # 1644487-2020-00214 captures the high impedance seen with programmer 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Common Device Name
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
MDR Report Key9687833
Report Number1644487-2020-00213
Device Sequence Number19701663
Product Code LYJ
UDI-Device Identifier05425025750429
UDI-Public05425025750429
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODEL 3000
Was Device Available for Evaluation? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2020
Initial Report FDA Received Date02/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age27 YR
-
-