Reported event: an event regarding instability involving an triathlon insert was reported.The event was confirmed medical review.Method & results: -device evaluation and results: visual inspection - visual inspection was performed as part of the material analysis report (mar), dated which indicated that - the returned tibial insert is shown.Damage was observed on the distal surface of the insert which is consistent with the explantation process.Backside impression markings on the distal surface of insert are consistent with contact against the tibial base plate.Damage was observed on the articulating surface of the insert which is consistent with contact against the femoral component.The damage present on the articulating surface is consistent with burnishing and scratching; these are common damage modes of uhmwpe.Damage was observed on the anterior surface of the insert which is consistent with the explantation process.Damage consistent with material loss due to delamination was observed on the posterior aspect of the medial condyle.Ma - the damage on the articulating surface of the insert was consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.Damage consistent with material loss due to delamination was observed on the posterior portion of the medial condyle.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Functional and dimensional inspection not performed as device was received in damage condition.-clinician review: : a review of the provided x-rays by a clinical consultant rejected and stated that : no clinical or pmh, no examination of explanted components.Based upon the information available for review, confirmation of instability of the left tka is noted on x-ray, but no determination can be made regarding the cause of this clinical event." -device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: it was reported that patient knee was revised due to instability.The event was confirmed by medical review.Visual inspection - visual inspection was performed as part of the material analysis report (mar), dated which indicated that - the returned tibial insert is shown.Damage was observed on the distal surface of the insert which is consistent with the explantation process.Backside impression markings on the distal surface of insert are consistent with contact against the tibial base plate.Damage was observed on the articulating surface of the insert which is consistent with contact against the femoral component.The damage present on the articulating surface is consistent with burnishing and scratching; these are common damage modes of uhmwpe.Damage was observed on the anterior surface of the insert which is consistent with the explantation process.Damage consistent with material loss due to delamination was observed on the posterior aspect of the medial condyle.Ma - the damage on the articulating surface of the insert was consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.Damage consistent with material loss due to delamination was observed on the posterior portion of the medial condyle.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.A review of the provided x-rays by a clinical consultant rejected and stated that : no clinical or pmh, no examination of explanted components.Based upon the information available for review, confirmation of instability of the left tka is noted on x-ray, but no determination can be made regarding the cause of this clinical event." the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|