| Catalog Number |
0380544100 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device came off during the procedure which led to a 30 minute delay.The piece was retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device came off during the procedure which led to a 30 minute delay.The piece was retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: inner workings of the shaverblade broke off.Probable root cause: deformed cutter teeth, incorrect gap between cutter/housing no/minimal clearance between cutter and housing , excessive runout, excessive wear/gall, incorrect diamond grit size/ type.Inadequate coating process.The reported failure mode will be monitored for future reoccurrence.
|
| |
|