Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes. a review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Device evaluation: analysis of the returned device identified: underweight and broken on anterior.A visual and microscopic analysis was performed which identified: striated broken, missing shell (0-25%) and crease fold.Based on the device analysis the final assessment is: a striated broken due to surgical damage consist in the use of some surgical tool.Missing shell due to inconclusive.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and capsular contracture, baker grade unknown.
|