ALLERGAN (COSTA RICA) STYLE 45 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 45-600
Medical Device Problem Codes Material Rupture (1546); Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 08/28/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.  a review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Device evaluation: analysis of the returned device identified: underweight and broken on anterior.A visual and microscopic analysis was performed which identified: striated broken, missing shell (0-25%) and crease fold.Based on the device analysis the final assessment is: a striated broken due to surgical damage consist in the use of some surgical tool.Missing shell due to inconclusive.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and capsular contracture, baker grade unknown.
 
Event or Problem Description
Healthcare professional reported right side capsular contracture, baker grade unknown and rupture.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 45 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9688959
Report Number9617229-2020-02104
Device Sequence Number16992445
Product Code FTR
UDI-Device Identifier10888628002111
UDI-Public10888628002111
Combination Product (Y/N)N
Initial Reporter StateIL
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/16/2020
Device Model Number45-600
Device Catalogue Number45-600
Device Lot Number2728779
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2019
Initial Report FDA Received Date02/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age33 YR
-
-