| Model Number |
CC4204A |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Brand name: collamer ultraviolet-absorbing posterior chamber single piece foldable intraocular lens.No similar complaint type events reported for units within the same lot.The lens was returned in liquid, in a lens vial.Visual inspection found that the haptic and optic were torn.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter states that before the surgeon injected a cc4204a intraocular lens, diopter +18.5, he "looked at the lens in the injector and noticed that something looked wrong so he ejected the lens outside the eye and the lens was torn." there was no patient contact with the lens.This occurred on (b)(6) 2019.
|
| |
|
Additional Manufacturer Narrative
|
|
H6-previous statement-lens work order search: no similar complaint type events reported for units within the same lot.H6-corrected statement-lens work order search: two similar complaint type events reported for units within the same lot.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h3- "delivery system was returned in lens box.Visual inspection found the plunger bent." should be in the initial mdr.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|