MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL WITH PUREFIX HA 62MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL WITH PUREFIX HA 62MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 542-11-62H
Medical Device Problem Codes Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Codes Extraskeletal ossification (1428); Pain (1994); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Inadequate Osseointegration (2646)
Date of Event 12/28/2008
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Not returned.
 
Event or Problem Description
Patient called stating he had a right hip replacement done on (b)(6) 2008 and is experiencing pain as he walks.Patient also reported that his doctor mentioned bone loss and loosening in his hip.Patient would like to know if his implants were part of a recall.
 
Event or Problem Description
Patient called stating he had a right hip replacement done on (b)(6) 2008 and is experiencing pain as he walks.Patient also reported that his doctor mentioned bone loss and loosening in his hip.Patient would like to know if his implants were part of a recall.
 
Additional Manufacturer Narrative
An event regarding loosening & pain involving a trident shell was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as product remains implanted.- clinician review: a review of the provided medical records and x-rays by a clinical consultant indicated: x-ray printouts include a series dated november 1, 2019, which is an ap of the pelvis and lateral of the right hip demonstrating bilateral uncemented total hip arthroplasties with no screws in the acetabular component.The right hip demonstrates brooker iii heterotopic ossification.X-rays dated (b)(6) 2019 are an ap of the pelvis and lateral of the right hip, which is unchanged.Based upon the information available for review, there is no indication that any symptoms described as related to the right total hip arthroplasty are due to the prosthetic components and the recall of the implanted prosthetic femoral head.Persistent lumbar pathology and radicular pain, as well as heterotopic ossification are more likely the cause of the symptoms described.-device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: the patient would like to know if his implants were part of a recall.The trident shell was determined not to be involved in recall.The exact cause of the event could not be determined because insufficient information was provided.Additional information including return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving a trident shell was reported.The event was confirmed.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned devices was performed as part of the material analysis: "images depicted show the accolade stem, the lfit head, and the trident insert assembled in the trident acetabular shell.On visual examination, bone growth was observed on the coated region of the stem.Yellow discolouration on the trident insert consistent with oxidation, possibly a result of synovial fluid absorption.Mechanical damage was also observed on the insert.Stereo microscope image shows the presence of discolouration and material wear on the stem trunnion.Explantation damage which was observed on all sides of the stem.Dark discolouration observed within the head taper with figure 6 showing black discolouration and material wear on the side walls of the head taper.A stereo microscope image of marks observed on the proximal surface of the head." -clinician review: a review of the provided medical records by a clinical consultant indicated: "on 7/3/23 the tha in question was revised for pain, elevated serum metal ions and suspicion of component loosening.Intra-operatively metalosis was described with black staining of the tissues.The acetabulum was not well fixed and easily removed.Upon removal of the metal head black material was seen on the trunnion but the trunnion itself was "intact" heterotopic ossification was also noted.It is confirmed that the patient underwent a revision tha for pain, acetabular lack of fixation, metallosis and elevated serum metal ions.The root cause of the lack of acetabular fixation and trunnionosis cannot be determined from the documentation provided." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to pain, loosening of the acetabular component, and trunnionosis/metallosis at the stem-head junction.The event was confirmed via clinician review of the provided medical records.The root cause of loosening of the acetabular component could not be determined from the information provided.However, it should be noted that the cocr head has been identified to be within scope of nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient called stating he had a right hip replacement done on (b)(6) 2008 and is experiencing pain as he walks.Patient also reported that his doctor mentioned bone loss and loosening in his hip.Patient would like to know if his implants were part of a recall.Update per the medical records received: the patient was revised on (b)(6) 2023 due to metallosis/trunnionosis at the stem-head junction and loosening of the acetabular component.The entire construct was revised to competitor implants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT PSL WITH PUREFIX HA 62MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key9693264
Report Number0002249697-2020-00258
Device Sequence Number16571231
Product Code LPH
UDI-Device Identifier04546540240545
UDI-Public04546540240545
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Model Number542-11-62H
Device Catalogue Number542-11-62H
Device Lot Number9RVMHE
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 01/14/2020
Supplement Date Received by Manufacturer02/07/2020
08/26/2024
Initial Report FDA Received Date02/11/2020
Supplement Report FDA Received Date03/06/2020
09/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/23/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization; Required Intervention;
Patient Age59 YR
Patient SexMale
-
-