| Model Number |
47127 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/24/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The returned implants are currently being evaluated.A follow up report with results of the investigation will be submitted upon completion.
|
| |
|
Event or Problem Description
|
|
Existing zodiac construct of the l4-s1 was extended one level (l3) with the arsenal fixation system on (b)(6) 2019.1 week following the fusion, the patient heard a pop sound while stretching his back.X-rays were ordered found the left l3 setscrew had disassociated itself from the l3 pedicle screw.Revision surgery took place on (b)(6) 2019 to remove and replace both the arsenal set screw and pedicle screw at the l3 with the zodiac.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of the set screw showed that the threads are not deformed which indicates that set screw was not cross threaded.However, the set screw bottom surface showed unusual wear marks.The bottom surface of the set screw is the surface that interacts with the rod to create a stable construct, therefore the wear pattern on this surface may be indicative of the screw performance.The explant set screw does not exhibit the hour glass wear mark.This indicates a non-distributed or normalized loading condition at lock down.Furthermore, the set screw showed wear marks similar to a starburst pattern which indicates set screw back out due to repetitive loading.This starburst pattern being indicative of a non-normal loading of the rod under a dynamic loading conditions.Based on these findings, the explants indicate that set screw disengagement is a result of a non-distributed loading and non-normalized rod conditions at final lockdown.
|
| |
|
Search Alerts/Recalls
|