The engineering evaluation stated: for psts event 43883, it was reported that a 6m02 device frayed during a procedure.No device was returned.Because the device was discarded at the hospital, it was not possible to perform a physical analysis of the device that would include images.Existing controls to prevent occurance of reported issue: 100% of devices undergo visual inspection at cut thread inspection and a tactile inspection prior to loading.If the requirements of these inspections were not met the product would have been rejected.Risk documents review: the suture hazards and harms analysis, design fmea, and process fmea already address the failure mode and hazardous situations resulting from frayed devices.Other tasks for the event: manufacturing records were reviewed and indicated that the reported lot met all pre-release specifications.Additionally, manufacturing records were reviewed for the associated cut thread lot and indicated that the lot met all pre-release specifications.No ncr or other quality objects were associated with the lot.Root cause: based on the investigation, the root cause of the frayed suture is not able to be determined.
|