• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ON-X LIFE TECHNOLOGIES, INC. ON-X PROSTHETIC MITRAL VALVE WITH STANDARD SEWING RING; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ON-X LIFE TECHNOLOGIES, INC. ON-X PROSTHETIC MITRAL VALVE WITH STANDARD SEWING RING; HEART-VALVE, MECHANICAL Back to Search Results
Model Number ONXM-31/33
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Death (1802)
Date of Event 01/14/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
A review of the available information was performed.Onxm-31/33 sn (b)(4) was implanted in the mitral position of a (b)(6)-year-old female patient.Date of surgery was (b)(6) 2019.Transesophageal echo (tee) at that time indicated a mean pressure gradient of 3mmhg, no insufficiency, and no leak.On (b)(6) 2020 (222 days post-implant), the patient was exercising, felt poorly, and reported to the emergency room.Echocardiogram indicated one of the two valve leaflets was not present.Re-operation attempted to replace the on-x valve with a tissue valve.Patient suffered cardiogenic shock and metabolic acidosis, developed prolonged hypertension, hypo-perfusion, followed by multiorgan failure before expiring.Date of death was not specified but likely occurred on or near the date of the reoperation.A photograph of the explanted mechanical valve shows a missing leaflet, but no other obvious irregularities.Radiographic scans of the patient show a large dislodged section of valve leaflet apparently wedged within the abdominal aorta at about the level of the l2 vertebra.The imaged leaflet was not available for inspection, but limited records seem to suggest it was not extracted from the patient.Furthermore, close inspection of the radiographic imaged leaflet clearly shows a significant portion of the dislodged leaflet is not visualized, indicating a fractured component.The location of the non-visualized portion of leaflet is not known, however, it is recognized as a section of the leaflet that normally helps keep the leaflet engaged with the valve housing.Cryolife has reviewed the manufacturing records and it was confirmed that all dimensional and performance specifications were met.Multiple requests have been made to the hospital to obtain additional information and for return of the explanted valve, or portions of the explanted valve, for further evaluation.To date the hospital has stated they will not be providing further information including operative notes, other medical records, autopsy findings, or return the explanted valve, or its components, for review.Additionally, the hospital will not allow review on-site by cryolife personnel under supervision.Consequently, the potential cause of the fracture is presently undetermined.Should additional information become available it will be evaluated.Based upon review of the limited information, this appears to be a case of structural valve dysfunction, namely a broken leaflet, of undetermined etiology.The instructions for use [ifu] list prosthesis structural dysfunction as a potential complication of prosthetic valve implantation with the possibility of explantation, reoperation, and/or death.
 
Event or Problem Description
According to the initial report, the patient was implanted with onxm-31/33 in the mitral position on (b)(6) 2019.On (b)(6) 2020 an echocardiogram showed a leaflet was missing and had migrated.This resulted in cardiogenic shock and metabolic acidosis.The surgeon attempted mitral valve replacement, but was unsuccessful.The patient developed prolonged hypertension and hypo perfusion, followed by multi-organ failure.The patient did not live.Radiographic scans show the dislodged valve at the level of the l2 vertebra.Reference medwatch report mw5092348.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-X PROSTHETIC MITRAL VALVE WITH STANDARD SEWING RING
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ON-X LIFE TECHNOLOGIES, INC.
1300 e. anderson ln., bldg. b
austin TX 78752
MDR Report Key9694503
Report Number1649833-2020-00089
Device Sequence Number1809906
Product Code LWQ
UDI-Device Identifier00851788001297
UDI-Public00851788001297
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P000037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/19/2024
Device Model NumberONXM-31/33
Was Device Available for Evaluation? No
Distributor Facility Aware Date01/14/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2020
Initial Report FDA Received Date02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Hospitalization;
Patient Age50 YR
-
-