• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ZERO TIP; DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ZERO TIP; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number M0063901030
Medical Device Problem Codes Break (1069); Deformation Due to Compressive Stress (2889); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medwatch #: mw5090836.(b)(4).Visual inspection found the basket was closed when received.The sheath is bent, buckled, and detached in the distal section.Functional inspection found the basket could not be open due to the damages observed on the sheath.Based on all available information, the investigation concluded that the failures observed are issues that could have been generated due to handling or manipulation of the device during the procedure, leading to the reported failure.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.
 
Event or Problem Description
It was reported to boston scientific corporation that a zero tip retrieval basket was used in a procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the basket failed to open.There were no patient complications as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.Investigation results revealed that the sheath is torn at distal end; therefore, this is now an mdr reportable event.
 
Event or Problem Description
It was reported to boston scientific corporation that a zero tip retrieval basket was used in a procedure performed on (b)(6), 2019.According to the complainant, during the procedure, the basket failed to open.There were no patient complications as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.Investigation results revealed that the sheath is torn at distal end; therefore, this is now an mdr reportable event.Please see block h10 for full investigation details.
 
Additional Manufacturer Narrative
Medwatch #: mw5090836.Block h6: device code 1069 captured the reportable investigation finding of sheath detached at distal section.However, upon reevaluation of the product investigation, it was discovered the sheath was not detached or torn.Therefore, this no longer constitutes a reportable event.Block h10: visual inspection found the basket was closed when received.The sheath is bent and buckled in the distal section.Functional inspection found the basket could not be open due to the damages observed on the sheath.Based on all available information, the investigation concluded that the failures observed are issues that could have been generated due to handling or manipulation of the device during the procedure, leading to the reported failure.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZERO TIP
Common Device Name
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key9695141
Report Number3005099803-2020-00360
Device Sequence Number13888876
Product Code FFL
UDI-Device Identifier08714729342021
UDI-Public08714729342021
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2020
Device Model NumberM0063901030
Device Catalogue Number390-103
Device Lot Number0023724425
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2019
Initial Date Received by Manufacturer 01/16/2020
Supplement Date Received by Manufacturer02/13/2020
Initial Report FDA Received Date02/11/2020
Supplement Report FDA Received Date03/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-