| Catalog Number |
07P92-30 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.There is no further patient information provided due to privacy issues.This report is being filed on an international product, list number 7p92 that has a similar product distributed in the us, list number 6c06.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated alinity i total psa results on one patient.The results provided were: on (b)(6) 2019 sid (b)(4) = 4.96ug/l (diagnostic range 0.0 - 3.9ug/l).The patient is being monitored post prostate cancer treatment and his previous results were less than 0.01ug/l.Repeat sample on (b)(6) 2020 = less than 0.01ug/l.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of complaint tickets for the assay nor testing could be completed as the lot is unknown.A review of complaints determined there are no trends for the product for the complaint issue.Return testing was not completed as returns were not available.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation no product deficiency was identified for the alinity i total psa assay.
|
| |
|
Additional Manufacturer Narrative
|
|
Suspect medical device, 4.Catalog# from 07p92-31 to 07p92-30; udi from (b)(4) to (b)(4).
|
| |
|