A (b)(6) advia centaur xp hcv (ahcv) result was obtained for a patient sample.On january 22nd, the hcv-rna testing was performed and the result was (b)(6).The patient sample was tested again on the same advia centaur xp system and the result was (b)(6).The patient sample was tested on an alternate method on the same day at a different location.The results were (b)(6).On january 23rd, the patient sample was tested at another laboratory and the result was (b)(6) on the alternate method.On january 24th, the patient sample was tested at another laboratory on the advia centaur xp.The hcv result was (b)(6).The initial results were reported to the physician.It is unknown if the results were questioned.Patient treatment was not altered or prescribed.There was no report of adverse health consequences due to the discordant advia centaur xp hcv result.
|
Siemens filed the initial mdr 1219913-2020-00036 on february 11, 2020.March 10, 2020 additional information: siemens has completed the investigation into the false negative hcv result.There was no quality control (qc) run on the day the samples were run.However, the sample was tested at an alternate lab and lot number and repeated negative.Multiple attempts were made to gather more information to investigate for potential root cause, but no response was received.There has been no further reports of discordant samples from this customer.It is possible that the patient had a very old infection and was recovered or it is possible that this was a case of a false positive result on the alternate method.It had been indicated that pcr was run however no data was submitted.Hcv lot in house data was reviewed and all patient pools, medical decision pool and quality control were resulting as intended.No product non conformance was identified.The instrument is performing within specifications.No further evaluation of the device is required.Mdr 1219913-2020-00037 supplemental report 1 and 1219913-2020-00038 supplemental report 1 were filed for the same event.
|